How to Apply NPWT (continued)
8. Canisters should be replaced as required or weekly. To change canister, make sure
power unit is in stand-by mode (if still running, press and hold the UNLOCK button,
followed by the RUN/STOP button). Clamp Wound Care Set tubing and remove by
turning anticlockwise and lifting out of tubing port (this can be reconnected to
new canister and unclamped if wound dressing is not being changed). Remove
sealing plug from its location on top corner of canister and use to cap tubing port
to seal in contents. Rotate locking knob 1/4 turn anticlockwise and remove canister.
Dispose of used canister according to local clinical waste policy. If continuing NPWT,
attach new canister and connect Wound Care Set tubing as previously described.
9. To stop the power unit, press and hold the UNLOCK button until power unit beeps
and L’ clears from display screen. Then press and hold the RUN/STOP button until
power unit beeps three times to return to stand-by mode (if using the power adapter,
disconnect it from the power unit and unplug from the power source). The power unit
will then power off automatically after one minute of inactivity or may be forced off by
pressing the MUTE key.
10. Place the user manual in a safe place for future use.
User Information
lThe VENTURI
®
AVANTI
/
VENTURI
®
COMPACT power unit will automatically lock 1 minute
after last button operation when running to prevent inadvertent operation of button
functions (except MUTE), as indicated by 'L' on the display screen. Press and hold the
UNLOCK button until power unit beeps if further button operation is needed (i.e. change
of therapy mode or vacuum level, or returning power unit to stand-by mode). The power
unit will lock again 1 minute after last button operation.
Please note that when running, the VENTURI
®
AVANTI
/
VENTURI
®
COMPACT power unit
must be unlocked before it can be returned to stand-by mode, i.e. the power unit cannot
be put into stand-by mode when the ‘L’ is displayed on the screen.
lDuring operation, the power unit should be placed upright on a horizontal surface with
the display uppermost. The VENTURI
®
AVANTI power unit can be placed in the optional
metal holder for use at the bedside or attached to an IV pole using the optional bracket.
The VENTURI
®
COMPACT is supplied with a bracket for IV pole attachment. To allow
patient full mobility during therapy, both power units can also be placed in the carry
cases provided. In all instances the power unit/canister assembly should be kept upright
during use to ensure correct operation. The VENTURI
®
COMPACT will display a warning if
the unit is over tilted.